Hydrocarbon resistant glass oxygen sensor patch for non-invasive monitoring of dissolved oxygen in Hydrocarbons, Organic solvents, fuels, and oils. The Outer Diameter of the patch is 0.25" (6.35 mm) and its thickness is 0.03" (0.76 mm). Pack of 5.

They are connected to our reflection probe and DOPO2 for oxygen measurement.

 

The Outer Diameter of the patch is 0.25" (6.35 mm) and its thickness is 0.03" (0.76 mm), with a response time (t90) less than 1 Sec.

Custom diameter available per customer request. These patches are pre-calibrated with a known calibration coefficient and slope (using DOPO2 EOM). Pack of 5


 

HCR formulation (sol-gel) Specifications

Response rate of patches

T90 < 1 Sec
T90~1 Sec with Black Overcoat

Dynamic range in gas

0-100% O2/0-760 mmHg O2

Dynamic range in water (DO)

0-40 ppm (wt)

Dynamic range in other liquids

0-oxygen saturation level

Resolution in gas

0.05% O2 / 0.4 mmHg

Resolution in water (DO)

0.02 ppm (wt) / 20 ppb (wt)

Resolution in other liquids

0.05% of oxygen saturation level

Accuracy

5% of reading

Lowest detectable level (gas)

0.05 % O2 / 0.4 mmHg

Lowest detectable level (water)

0.1 ppm (wt) / 100 ppb (wt)

Lowest detectable level in Hydrocarbons

0.25% of oxygen saturation level

Sensor drift

0.001 O2% per hour

Operating temperature

-50 to + 120

Recommended patch replacement

1 year

 